ELOTEST IS500 Box
The 2-channel eddy current test instrument is characterized by a fully digital signal processing. It allows a single crack inspection as well as a multi-frequency investigation of the materials structure, but also a combination of both applications. The test instrument can be integrated directly into a production line. The process communication can be handled via a universal digital I/O interface or via serial FieldBus Interface. In addition, it offers numerous functionalities such as electronic distance compensation, multi-frequency multiplexing, multi-sensor multiplexing and multi-frequency/multi-lot sorting with self-learning "Bubble Gates".
Zurück zur Übersicht Data Sheet ELOTEST IS500 BoxFeatures
- Fully Digital Signal Processing Chain
- Powerful Multiplexing
- Powerful Sorting Capabilities
- Brilliant Signal Display
- Integrated I/O-Interfaces
- high dynamic range (96 dB)
- frequency range: 10 Hz – 12 MHz
- band width: 10 kHz
- ultra-fast multiplexing (multiplex rate up to 32 kHz)
- independently adjustable high- and low-pass filters (1 Hz up to 100 kHz) incl. band pass
- 4 thresholds for each MUX channel available
- parameter-multiplexing (max. 8 parameter sets), e.g. for sequential switching of test frequency, amplifier settings, phase setting or filter frequencies
- sensor-multiplexing for up to 8 sensors with external multiplexer module (option)
- multiplex rate depending on the selected test frequency up to 32 kHz

- ultra-fast 1-frequency sorting (with reversal point detection)
- special sorting module with up to 8 test frequencies (multiplexed, 10 Hz – 150 kHz) for up to 8 “GOOD”-lots (“MultiLot”), guided teaching of “GOOD”-parts, retroactive teaching of “GOOD”-parts (“Retro-Teach”) and self-learning “Bubble Gates”

- 9″ WVGA colour display (800 x 480 Pixels, 16:9 format)
- simultaneous display of up to 8 signals with a display rate of 250,000 signal dots per second for each channel (in real time)
- adjustable fading time
- 2 sensor connectors (26-pin)
- analogue output (eddy current data, max. + 10 V)
- universal-I/O-module (16 dig. inputs and 24 dig. outputs (24 V DC, optocoupled) and 2 counter inputs (for incremental encoders)
- programmable logic for sorting switches, control signals for mechanical sorting units
- diverse fieldbus modules, e.g. for ProFiBus, ProFiNet, EtherCat (optional)